Determination of relative and absolute efficiency functions in the range of 122 keV ÷ 8.5 MeV of HPGe detector

Son An Nguyen 1, *
Lanh Dang 1
Minh Van Truong 2
  1. Dalat University
  2. Dong Nai University
Correspondence to: Son An Nguyen, Dalat University. Email: pvphuc@vnuhcm.edu.vn.
Volume & Issue: Vol. 18 No. 2 (2015) | Page No.: 79-87 | DOI: 10.32508/stdj.v18i2.1135
Published: 2015-06-30

Online metrics


Statistics from the website

  • Abstract Views: 2567
  • Galley Views: 1150

Statistics from Dimensions

Copyright The Author(s) 2023. This article is published with open access by Vietnam National University, Ho Chi Minh city, Vietnam. This article is distributed under the terms of the Creative Commons Attribution License (CC-BY 4.0) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. 

Abstract

Construction of detector is necessary. However, on large energy range the manufacturers could not also support the explicit function of relative and absolute efficiencies of detectors. One of the reasons is a restriction of energy range of gamma sources (normally < 3 MeV). This paper presents the results of construction of relative and absolute efficiency functions within a range from 122 keV to 8.5 MeV. The sources are used combining 152Eu point source and 36Cl activated isotope by thermal neutron captured reaction 35Cl of Dalat nuclear reactor (DNR) by 35Cl(n, γ)36Cl reaction. This result can be applied in determining quantitative analysis of samples of neutron activation and radioactivity chemistry