Downloads
Abstract
The advantage of gamma – gamma coincidence method is abilities of background reduction. This technique has been studied and applications for neutron activation analysis. The experimental neutron activation analysis of geological samples, biological samples and environmental samples has been conducted by several some laboratories in the world. In this study, we would like to present the results of Arsenic and Selenium determination in geology sample by neutron activation analysis with technique of gamma – gamma coincidence measurement. By application of this gamma – gamma coincidence method, the influence isotopes as 182Ta, 181Hf, 152Eu were eliminated output from spectrometric measurement when analyzing 75Se and effects of overlap peaks from spectrometric measurement when analyzing arsenic.Thus improved detection limits for Arsenic and selenium in geological samples when analyzed by neutron activation analysis method.
Issue: Vol 17 No 2 (2014)
Page No.: 82-89
Published: Jun 30, 2014
Section: Natural Sciences - Research article
DOI: https://doi.org/10.32508/stdj.v17i2.1317
Download PDF = 904 times
Total = 904 times